Contamination Monitor CoMo 170 |
The CoMo 170 is an efficient, portable contamination monitor
equipped with an innovative detector technology for the high sensitive and nuclide
referred measurement of surfaces regarding alpha, beta and gamma contaminations.
The detector technology with thin plastic scintillation detector completely works without gas-filled or gas flow proportional detectors, thus avoiding expensive costs for consumable gas or for repair of Xenon detectors.
Measurement of alpha, beta and gamma radiation with only one detector. The measuring system automatically identifies the existence of alpha radiation. It is possible to effect a simultaneous, selective measurement of alpha and beta/gamma contaminations.
A defective detector foil can be replaced by the user himself. An expensive repair by the manufacturer is not necessary.
The nuclide referred efficiencies of the
scintillation detector can be compared with the efficiencies of common counter
tubes.
Features: |
Technical Data: |
| Detector type: | thin-layer plastic scintillation detector with ZnS coating |
| Detector size: | 170 cm² (mechanically protected by a honeycombed protective grid) |
| Background count rate: | alpha: approx. 0,1 cps
beta/gamma: approx. 15-25 cps |
| Background count rate subtraction: | with adjustable measuring time |
| Alarm: | for each nuclide separately adjustable, acoustic warning |
| Display of measured values: | either in cps or nuclide referred in Bq or Bq/cm²
digital and analog (bar graph) display |
| Nuclides: | 25 nuclides, pre-set calibration factors, variable acc. to user requirements (user-specific nuclides can be added), integrated auto-calibration |
| Measuring time: | continuous (adjustable attenuation), for stationary operation adjustable in sec |
| Display: | large, graphic LC-display (128x64 pixels), with illumination (illumination time adjustable) |
| Temperature range: | -10 °C up to +40 °C, special version up to -20 °C |
| Power supply: | 2 batteries, AA Mignon or corresponding accumulators (NiCd, NiMH), rechargeable by recharger unit or by wall station during stationary use (option) |
| Dimensions:
Weight: |
(280 x 125 x 135) mm / length including handle
approx. 750 g (batteries included) |
| Interfaces: | |
| PDF-file: | CoMo170_en.pdf (143 KB) |
Efficiencies of various radionuclides(average values of measurements with 100 cm² substances) |
The nuclide referred efficiencies of the scintillation detector are comparable with those of common GM-tubes.
Special edition: CoMo 170Ffor nuclear operations at fire brigades |
Different technical data: |
| back to top |