Welcome
| Company Profile
| Products
| GRAETZ News
| Service >
Functional Tests
> GRAETZ Dose Rate Alarm Units
| Downloads
| Contact
| Imprint
|
- Visual control regarding damages on the outside of the instrument.
- Insert the batteries (observe right polarity!). The green battery control LED
flashes, indicating the batteries are OK. Otherwise, the batteries have to be replaced.
- Check the GM-tube under the influence of ambient background radiation: within
2 minutes approx. 1 count has to be triggered. The red LED flashes and an acoustic
alarm sounds. Due to the statistically varying dose rate of ambient background
radiation the background counts are spread irregularly, too.
If a defect is observed during the control measurements, which cannot be cleared
by yourself, the instrument has to be returned to the manufacturer
GRAETZ for repair.
- Visual control regarding damages on the outside of the instrument.
- The green battery control LED flashes, indicating the accumulator is OK.
Otherwise, the accumulator has to be recharged by means of the recharger unit
GRAETZ "462 L".
- Check the GM-tube under the influence of ambient background radiation: within
2 minutes approx. 1 count has to be triggered. The red LED flashes and an acoustic
alarm sounds. Due to the statistically varying dose rate of ambient background
radiation the background counts are spread irregularly, too.
If a defect is observed during the control measurements, which cannot be cleared
by yourself, the instrument has to be returned to the manufacturer
GRAETZ for repair.
- Visual control regarding damages on the outside of the instrument.
- Control the battery contacts in the battery compartment. They have to be
non-corrosive and dry.
- Visual check of the membrane in the sound transmission aperture. It may not
be damaged.
- Insert the batteries (observe right polarity!). Turn the switch to
"TEST". The instrument must trigger a permanent signal, the LED has to light up.
In case of an alarm sound and a flashing LED, the batteries have to be replaced.
- Turn the switch to alarm threshold 25 µSv/h. The LED will flash in irregular
intervals being a permanent control of the detector, the electronics and the battery
voltage.
- For additional measurements exceeding ambient background radiation a radioactive
test source (e. g. GRAETZ "PS 9") is needed. Generally, 137Cs with an
activity of 333 kBq is used. The half life of Caesium is 30 years. Remove the test
source from the lead storage container and hold it onto the point shaped marking
(= position of the GM-tube). When perfectly working, the alarm thresholds 25 µSv/h
and 40 µSv/h will trigger an alarm signal. After the control measurements the test
source has to be put into the storage container again.
If a defect is observed during the control measurements, which cannot be cleared
by yourself, the instrument has to be returned to the manufacturer
GRAETZ for repair.
- Visual control regarding damages on the outside of the instrument.
- Control the plug connection. It has to be non-corrosive and dry.
- Control the battery contacts in the battery compartment. They have to be
non-corrosive and dry.
- Use a set-screw to untwist the knob of the "ON"-switch. Remove the spring ring
and the cover. Take care that the membrane over the sound transmission aperture is
not damaged. After the test cover, spring ring and knob have to be mounted again.
- Insert the batteries (observe right polarity!) and set switch to "TEST".
The instrument must trigger an alarm signal. The LED must not light up. If the LED
lights up and no alarm signal is triggered, the batteries have to be replaced.
- For additional tests a radioactive test source (e. g. GRAETZ PS 9) is needed.
Generally, 137Cs with an activity of 333 kBq is used. The half life of
Caesium is 30 years. Remove the test source from the lead storage container and
put it onto the ring-like marking situated beside the fixing clip. When the device
is perfectly working, the alarm signal will be triggered when reaching the alarm
thresholds 25 and 50 µSv/h. After the control measurements the test source has to
be put into the storage container again.
If a defect is observed during the control measurements, which cannot be cleared
by yourself, the instrument has to be returned to the manufacturer
GRAETZ for repair.
- Visual control regarding damages on the outside of the instrument. If the
housing is damaged, don't put the alarm lamp into operation.
Otherwise:
- Set the knob for selecting the alarm thresholds to "Test". The green
battery control LED must light up at all adjusted alarm thresholds. If the LED
does not, the built-in accumulator has to be recharged (recharger unit is supplied
together with the GWL 10m).
- The functional test of the complete electronics will be started in the
switch position "Test", too. When the device is perfectly working, the alarm signal
will be triggered in the same way as if an alarm threshold is exceeded.
If the housing is damaged or a defect is observed during the control
measurements, which cannot be cleared by yourself, the instrument has to be
returned to the manufacturer GRAETZ for
repair.
Welcome
| Company Profile
| Products
| GRAETZ News
| Service >
Functional Tests
> GRAETZ Dose Rate Alarm Units
| Downloads
| Contact
| Imprint
|
[Copyright 2009 GRAETZ Strahlungsmeßtechnik GmbH]